B-34
D
ELTA
/D
ELTA
XL/K
APPA
VF8
B T
ECHNICAL
D
ATA
ECG/Arrhythmia/ST Supplemental information required by AAMI EC11/13 and
IEC 60601-2-27
Respiration excitation
waveform
Square wave signal, 50 µA, 39.896 KHz
Noise suppression
Not Applicable
Time to alarm for
tachycardia
Vent tachycardia 1mVpp, 206 bpm
Gain 0.5, range 3.0 to 3.5 seconds, average 3.3 seconds
Gain 1.0, range 2.9 to 3.3 seconds, average 3.2 seconds
Gain 2.0, range 2.8 to 3.5 seconds, average 3.0 seconds
Vent tachycardia 2 mVpp, 195 bpm
Gain 0.5, range 2.2 to 4.0 seconds, average 3.0 seconds
Gain 1.0, range 1.9 to 2.5 seconds, average 2.3 seconds
Gain 2.0, range 2.0 to 2.9 seconds, average 2.5 seconds
Tall T-wave rejection capa-
bility
Exceeds minimum 1.2 mV T-Wave amplitude required by ANSI/AAMI
EC 13 Sect. 3.1.2.1(c) and IEC 60601-2-27 Sect 6.8.2 bb) 2).
Heart rate averaging
method
Heart Rate is normally based on the average R-R interval calculated
over the last 10 seconds, however it updates more quickly to reflect
changes to the patient’s underlying rate.
Response time of heart rate
meter to change in heart
rate
HR change from 80 to 120 bpm:
Range: [3.4 to 7.1 seconds] Average: 5.3 seconds
HR change from 80 to 40 bpm:
Range: [6.3 to 8.6 sec] Average: 7.4 seconds
Heart rate meter accuracy
and response to irregular
rhythm
Ventricular bigeminy: 80bpm
Slow alternating ventricular bigeminy: 60 bpm
Rapid alternating ventricular bigeminy: 120 bpm
Bidirectional systoles: 90 bpm
Accuracy of input signal
reproduction
Methods A, B, C and D from AAMI EC11 were used to establish over-
all system error and frequency response.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...