ONITORING
Overview
When used with the EEG pod, the monitor measures up to four EEG signals. Each
EEG waveform has its own parameter box, and each parameter box displays up to
three selectable parameters. EEG parameters can be trended, printed, and displayed on
a central station.
Each EEG lead is a differential measurement between two electrodes (positive and
negative). For example, a lead label of Fp-Cz means the positive electrode is at the Fp
site and the negative electrode is at the Cz site.
Precautions
The following precautions apply to EEG monitoring during electrosurgery:
Place EEG electrodes above the mastoid bone.
Do not perform electrosurgery above the mastoid bone.
NOTE:
“EEG pod H/W Failure” message appears if the EEG pod software and the
monitor software are not compatible.
The EEG pod is intended for use with surface electrodes only.
The
Test Pulse
feature on the Biomed menu (see chapter 2) allows your
hospital’s technical personnel or DrägerService to confirm the accuracy of EEG
signals.
NOTE:
EEG parameters do not alarm.
EEG channel 1 is required for EEG monitoring to occur.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...