APPA
13-15
NBP Open Line
The inflation time during cuff inflation
cycle is too long or the inflation rate is
too low.
Check that the hose and cuff are
properly connected to the monitor.
NBP Out of
Range
The values are reported, but are out of
specified range.
No action is required.
NBP
Overpressure
The cuff pressure is over 270 mmHg
(Adult), 180 (Pediatric) or 140 mmHg
(Neonatal). The cuff deflates
automatically.
• Check the patient and treat, if
necessary.
• Check the cuff for obstructions.
• Retry the measurement.
NBP Retrying
The monitor failed to detect sufficient
pulsations, aborted the measurement,
then started a new one.
No action is required.
NBP
Overpressure
Circuit Failure
The cuff overpressure circuit has failed. Contact the hospital’s technical
personnel or Dräger DrägerService.
NBP Meas.
Circuit Failure
Software or hardware failure. A
persistent alarm of low priority sounds,
and NBP measurements are no longer
possible.
• Power cycle the monitor.
• Contact the hospital’s technical
personnel or DrägerService if the
error persists.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...