![Dräger Infinity Delta Instructions For Use Manual Download Page 448](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779448.webp)
26-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
26 P
ULSE
C
ONTOUR
C
ARDIAC
O
UTPUT
(P
I
CCO)
MONITORING
5. Fill the injectate temperature sensor housing, enclosed in the pressure
monitoring kit and connect it to the distal lumen of the central venous line.
6. Connect the p-CO thermistor cable to the p-CO intermediate cable.
7. Plug the p-CO intermediate cable into the port labeled “CO” on the PiCCO
pod.
8. Connect the “Injectate temperature sensor housing” to the p-CO thermistor
cable.
9. Slide the ART blood pressure transducer into a slot on the front of the PiCCO
pod. Make sure the slot’s pressure label is “ART”.
10. Slide the Central Venous Pressure transducer into another slot. Make sure this
slot pressure label is “CVP”.
11. Connect the PiCCO catheter with the blood pressure transducer.
12. Use the 10-pin cable to connect the blood pressure transducer to the Delta/
Delta XL/Kappa.
13. If you have not already done so, enter the patient specific input parameters
(height and weight) in the Patient Admit menu (see page 4-2).
NOTE:
PiCCO measurements are only possible after the arterial pressure transducer
is zeroed and the PiCCO pod is calibrated. For accurate measurements, calibrate the
PiCCO pod within five minutes of zeroing the pressure.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...