Cross-sensitivity Compensation Error
: Gas measurement accuracy was tested at the gas
levels listed in the following table. The additional worst case error when compensation is correctly
selected (for barometric pressure, O2, and N2O) for the actual fractional gas constituents that are
present is:
0 to 40 mmHg: ±2 mmHg additional error
41 to 70 mmHg: ±3 % additional error
71 to 99 mmHg: ±5 % additional error
Additional notes regarding cross-sensitivity compensation errors
:
Xenon and Helium: Constantly bias etCO2 values >= 38 mmHg by up to –5 mmHg
Desflurane: At concentrations > 5%, constantly biases etCO2 values >= 38 mmHg by up to
+4tommHg
Quantitative Effects Measured At
Gas or Vapor
Gas Level
(Gas levels in % volume fraction)
Nitrous oxide
60
Halothane
4
Enflurane
5
Isoflurane
5
Sevoflurane
5
Xenon
80
Helium
50
Metered dose inhaler
propellants
Not specified for use
Desflurane
15
Ethanol
0.1
Isopropanol
0.1
Acetone
0.1
Methane
1.0
Oxygen measurement
range:
5-100% O2
Accuracy:
One point calibration:
≤
14.2 % FS (at RTP)
Two point calibration:
≤
3 % FS (at RTP)
Warm up:
Value available immediately after calibration
Nominal response time:
97 % in 30 seconds (flow rate = 2L/min at RTP)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...