BIS S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
25-7
Show all
parameters
(continued)
PWR
: Total Power
(Range = 40 - 100 dB in the
frequency range from 0.5 -
30 Hz)
A measure of absolute total
power (in decibels) in the
frequency range from 0.5 -
30 Hz.
BCT:
Burst Count
(Range = 0 - 30)
An alternative to BSR for
quantifying the amount of
suppression in the EEG. It
gives the number of EEG
bursts in the last minute.
Burst Count is activated only
when an Extend sensor is
connected. BCT value is
blank for any other type of
sensor.
Check
impedance
Performs an impedance check (tests the
quality of BIS electrodes’ connection) of
the sensor (for more information, see
Not applicable
BIS smoothing
rate
Selects the smoothing rate at which BIS
value is averaged
•
15 sec.
(default): More responsive
to state changes (for example,
induction or awakening).
•
30 sec
: A smoother trend (lower
variability and less sensitive to
artifact).
Filter
Allows you to set filters to ON or OFF
NOTE
:
The BISx pod uses filters to screen out
undesirable interference from the raw
EEG signal. If you prefer to observe the
raw EEG without the filters, they may be
turned OFF.
•
ON
(default)
• OFF
Parameter 2
Allows you to set second BIS parameter
• BCT
• PWR
• SEF
•
BSR
(default)
SRS # of
Use(s)
available
Allows you to display the number (0-99) of
uses that remain available.
NOTE
:
This display is read only.
• Not applicable
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...