S
PECIALTY
M
ENUS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
2-15
Extend All Alarms
OFF
Determines whether clinical personnel can
use the
All Alarms Off
fixed key to extend
the
All Alarms Off
time.
NOTE
: This item only appears if your
monitor is configured for this feature
(contact your hospital’s technical
personnel for more information).
• Enabled
•
Disabled
(default)
Alarm
Validation
Validates the alarm conditions to limit
nuisance alarms due to artifact or motion/
movement by delaying time to alarm.
NOTE:
When the
Alarm Validation
setting is set to
ON
, the time to alarm from
the onset of a limit violation equals the time
for detection plus the designated alarm
validation signal delay. For HR, this time
may exceed the maximum of 10 seconds
as required by AAMI EC13 and IEC
60601-2-27.
Alarm validation times are:
Upper limitLower limit
HR
6 sec 6 sec
RESP
14 sec 14 sec
IBP
4 sec 10 sec
SpO
2
6 sec 10 sec
1
PLS
6 sec. 10 sec
1
SpO
2
*
4 sec 4 sec
PLS*
4 sec. immediately
All other parametersimmediately
1
See
SpO2 Alarm Delay
below.
•
ON
– Enables the alarm
validation.
•
OFF
(default) – Disables the
alarm validation.
SpO2 Alarm
Delay
Validates an SpO
2
alarm condition by
requiring that the violation persists for
10 seconds (lower limit) before sounding
an alarm.
NOTE
:
Alarm validation
must be turned
ON
. If
Alarm Validation
is turned
OFF
,
SpO2 Alarm Delay
is automatically
turned
OFF
and menu selection is
ghosted.
•
ON
– SpO
2
or PLS lower limit
alarm conditions are
annunciated after it persists for
a period of 10 seconds.
•
OFF
– SpO
2
or PLS lower limit
alarm condition is not validated
before annunciation.
ASY/VF Alarms
Allows user to prevent disabling of ASY
and VFIB alarms.
WARNING:
When the HR alarm
is set to OFF and ARR
monitoring is set to “OFF”, the
monitor cannot generate ASY/VF
alarms.
•
Always ON
– ASY and VFIB
alarms are always active.
•
Follow HR Alarms
(default) –
ASY and VFIB alarms follow the
HR alarm settings.
The Unit Manager Menu
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...