RESP marker
Superimposes a vertical line on the
RESP waveform when the monitor
detects a breath. If the monitor
displays the RESP marker because
of artifact or other interference, set
the breath detection threshold to
only count valid breaths:
1. Set the mode to
Manual
.
2. Enable
RESP Marker
.
3. Click on
Size.
4. Set the Size value at the lowest
value where the RESP marker
appears.
NOTE
:
• RESP markers are not transmitted
over the network and do not
appear on remote views or
recordings.
• The RESP marker indicates the
time of breath detection, not the
beginning or end of respiration.
• ON
• OFF
Coincidence
Detect
Identifies when the respiration rate
is within 20% of the heart rate,
indicating that the monitor may be
counting heart beats as respiration.
WARNING: Respiration/heart rate
coincidence could mask an
apnea condition.
NOTE
: Enable respiration alarms
before setting Coincidence Detect
to
ON
.
•
ON;
The
RESP Coincidence
message
appears whenever respiration/heart rate
coincidence is detected (default for
Neonatal patient monitoring).
•
OFF;
the monitor does not detect
respiration/heart rate coincidence (default
for adult and pediatric patient monitoring).
RESP
Monitoring
Enables and disables respiration
monitoring. This can also be done
from the Main Screen and ECG
setup menus (see pages 2-5 and 8-
18).
• ON
• OFF
Apnea Time
Sets the time the monitor waits
before reporting a cessation of
breathing as an apnea event and
sounds an alarm.
•
OFF
,
10 s
,
15 s
,
20 s
,
25 s
,
30 s
NOTE
: This feature is available in Neonatal
and Pediatric patient mode only.
Apnea
Archive
Allows user to store and/or record
automatically an apnea alarm
event. User can later review stored
alarms on Event Recall screen.
•
OFF
,
Record
,
Store
(default),
Str./Rec
.
NOTE
: This feature is available in Neonatal
and Pediatric patient mode only.
The Respiration Setup Menu
Menu Item
Description
Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...