![Dräger Infinity Delta Instructions For Use Manual Download Page 580](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779580.webp)
B-54
D
ELTA
/D
ELTA
XL/K
APPA
VF8
B T
ECHNICAL
D
ATA
MultiGas
Via Scio Module
Alarm limit range:
18 to100 %
Stability of accuracy:
±3 % over an 8-hour interval
Protection against electric
shock:
Type CF
Standards:
IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable sections
only, Oxygen analyzers for monitoring patient breathing mixtures); EN
ISO 21647:2009
Note
: RTP = Room temperature / pressure 23 °C±3 and ambient barometric pressure
Parameter display:
etCO2*, iCO2*, RRc*, et O2, iO2, O2, N2O, etHAL, iHAL,
etISO, iISO, etENF, iENF, etSEV, iSEV, etDES, iDES
NOTE:
Asterisks distinguish Scio module parameter labels
from those of the etCO
2
module.
Agents measured:
Halothane, Isoflurane, Enflurane, Sevoflurane,
Desflurane
NOTE:
With Scio Four plus and Scio Four Oxi Plus the
anesthetic agent identification threshold is max. 0.3 Vol%.
The agent identification threshold of a second agent is 0.4
Vol% or, if Desflurane is the primary agent, one tenth (10%)
of the Desflurane concentration.
Display values:
Agents: CO2, O2:Inspired and expired
concentrations (%), plus trend, and waveform
N2O:
Inspired and expired
concentrations (%)
Measurement methods:
CO2, Agents, N20:Infrared
O2:
Paramagnetic sensor
Display range:
CO2:
0 to 10 % (resolution 0.1 %)
O2:
0 to 100 % (resolution 1.0 %)
N2O:
0 to 100 % (resolution 1.0 %)
Agents:
Halothane:0 to 10 % (resolution 0.1 %)
Isoflurane:0 to 10 % (resolution 0.1 %)
Enflurane:0 to 10 % (resolution 0.1 %)
Sevoflurane:0 to 11% (resolution 0.1 %)
Desflurane:0 to 24 % (resolution 0.1 %)
Alarm limit resolution:
CO2: 0.1 %
O2, N2O: 1 %
Agents: 0.1 %
Full Accuracy after warm-up period complete
Agents:
1, 2, 3
(when RR
≤
60 with I:E of 1:2)
Halothane (up to 8.5 %vol):±(0.15 vol % + 15 % rel.)
Isoflurane (up to 8.5 %vol):±(0.15 vol % + 15 % rel.)
Enflurane (up to 10 %vol):±(0.15 vol % + 15 % rel.)
Sevoflurane (up to 10 %vol):±(0.15 vol % + 15 % rel.)
Desflurane (up to 20 %vol):±(0.15 vol % + 15 % rel.)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...