10-8
D
ELTA
/D
ELTA
XL/K
APPA
VF8
10 ST M
ONITORING
Changing the ISO and ST Points
When you change the ST and ISO measuring points on the Measuring Points screen,
the monitor recomputes the ST deviation value accordingly (see page 10-9 for
directions on changing the ST and ISO points). During this procedure, the changing
ST deviation values are displayed in yellow beneath the current values, which appear
in green.
At the bottom of the screen, the placement of the ISO measuring point (in
milliseconds) before QRS onset is displayed next to the label
‘ISO’
,
while the
placement of the ST measuring point (in milliseconds) after QRS offset is displayed
next to the label
ST
.
On all trend displays, markers indicate changes in the placement of measuring points
as well as Relearn operations. The labels
‘CHG’
(Change) and
‘LRN’
(Relearn) appear
in time-stamped columns in the Trend Table and in the ST Trend Graphs. Also in the
Trend Graphs, a solid white vertical line in the ST trend graphs marks the time of a
measuring point change, while a dotted vertical line marks the time a Relearn
operation was initiated. Use the cursor to pinpoint the time of marked changes and
Relearn operations (see page 6-3 for instructions on using the cursor in trend graphs).
NOTE:
It is good clinical practice to check the position of the isoelectric and ST
measuring points before starting ST monitoring.
After a Relearn is complete, the QRS onset and offset are locked until you
initiate another Relearn.
NOTE:
See page 9-4 for information about events and procedures that automatically
trigger a relearning period.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...