Arrhythmia Setup
Modes (Full, Basic, OFF)
As shown in the following table, the monitor reports certain arrhythmia events even if
you set
ARR Monitoring
to
OFF
.
Basic
arrhythmia mode allows you to expand the
list of events reported. When
ARR Monitoring
is set to
Full
, the monitor reports all
available arrhythmia events.
You can configure the monitor to process arrhythmia according to the number and
variety of events you wish to observe.
1. Click on the ECG parameter box to display the ECG menu.
2. Click on
ARR Monitoring
.
3. Scroll through the available settings (
Basic, Full
, or
OFF
)
and click to verify
your selection.
ARR Mode
Available Display Parameters
Parameter box
OFF
• ASY (asystole)
• VF (ventricular fibrillation)
• ARTF (artifact)
Basic
• ARR (label to register arrhythmia occurrence)
• ASY (asystole)
• VF (ventricular fibrillation)
• ARTF (artifact)
• VT (ventricular tachycardia)
Full
• All arrhythmia events (See page 9-3 for a
complete list.)
• PVC (premature ventricular contraction)
WARNING: When HR alarms are disabled and ARR monitoring
is turned off, the monitor cannot generate ASY/VF alarms. To
restore this ability, set ASY/VF alarms to “Always ON” (see
page 2-15).
NOTE:
If Full monitoring is installed as a locked option, you can select
Full
,
Basic,
or
OFF
. If not, the choices are
Basic
and
OFF
.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...