Overview
The monitor performs physiological calculations using data acquired by the monitor
and other devices. The monitor stores derived parameters and displays them on the
Results (Calculations) screen, Drug Calculator, or Labs display. Available inputs and
calculated parameters vary depending on whether you use standard or optional
calculations software.
As a standard feature, the monitor automatically calculates a set of hemodynamic
parameters called Hemo-Calcs whenever you measure cardiac output (see page 15-14
for detailed information). The monitor can also be configured to calculate drug-
related parameters, including concentration, rate, total dose, and total volume.
In addition to these standard calculation features, two additional features are available
with the PhysioCalcs software option:
Hemodynamics –
The monitor calculates hemodynamic parameters based on
cardiac output, invasive blood pressure and patient data (for example, height
and weight).
Hemo/Oxy/Vent Calculations
– This locked option provides oxygenation,
and ventilation parameters in addition to hemodynamic parameters (for a
complete list of hem/oxy/vent parameters, see page 16-5). When the monitor
is connected to WinView or a device using the Medical Information Bus
(MIB) interface protocol, you can also obtain laboratory data through the
Hemo/OxyVent menu.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...