Status Messages
Message
Possible Cause
Suggested Action
BT > UL
BT < LL
The blood temperature is outside the
alarm limits, due to one of the
following conditions:
• A physiological condition
• Inappropriate alarm limits
• A defective sensor or cartridge
• Check the patient and treat, if
necessary.
• Change the alarm limits.
• Check the equipment and replace
it, if necessary.
BT Out of Range
(High)
BT Out of Range
(Low)
The blood temperature is outside the
measurement range (25 °C to 43 °C),
because of a defective sensor or
cartridge.
Check the equipment and replace it,
if necessary.
C.O. Already In Use
The catheter and injectate probe are
connected to different hemo pods.
• Disconnect the intermediate
cables from the hemo pods.
• Connect both the catheter and the
injectate probe to the same cable
and reconnect them.
• Discard the unused intermediate
cable.
C.O. Out of Range
(High)
C.O. Out of Range
(Low)
The cardiac output value is greater
than 20 liters/min or less than 0.5 liter/
min because of the following
conditions:
• A physiological condition
• Unstable baseline
• Incorrect injectate volume, catheter
size, or Comp. Constant
• Defective catheter, cable, or
cartridge
• Check the patient and treat, if
necessary.
• Use cooler injectate.
• Enter the correct values in the
C.O. menu.
• Repeat the measurement. If the
message persists, replace the
defective components.
C.O. Injectate Too
Cold
The injectate is cooler than –5 °C.
Use an injectate within the range of
–5 °C to +30 °C.
Defective cable or hemodynamic pod. Check the equipment and replace it,
if necessary.
C.O. Injectate Too
Warm
The injectate is warmer than +30 °C.
Use an injectate within the range of
–5 °C to +30 °C.
The injectate probe is not connected.
Check probe connection. If problem
persists, replace the probe.
Defective cable or hemodynamic pod. Repeat the measurement. If the
problem persists, replace defective
part.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...