![Dräger Infinity Delta Instructions For Use Manual Download Page 229](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779229.webp)
NBP S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
13-11
Continuous Measurements
In continuous mode the monitor takes NBP measurements continuously over a period
of five minutes.
To start continuous measurements
1. Open the NBP menu (see page 13-8).
2. Click on
Continuous Mode
.
3. Click to toggle the mode
ON
. The monitor takes NBP measurements for five
minutes and continuously updates the NBP parameter box. The previous
measurement displays until the current one is complete.
The monitor waits at least two seconds between the end of one measurement and the
start of another to achieve reperfusion of the limb. The entire continuous measurement
cycle is aborted if there is an NBP alarm.
To stop a continuous measurement in progress, click again on
Continuous Mode
in
the NBP menu or press the
NBP Start/Stop
fixed key. The entire measurement cycle
is canceled.
WARNING:
Rapid, prolonged cycling of non-invasive pressure
measurements have on occasion been associated with
petechia, ischemia, purpura, or neuropathy. Make sure
that the cuff is properly attached and check the cuff
site regularly to prevent cuff pressure from impeding
the blood flow.
When using continuous mode, observe the patient
closely and verify limb perfusion clinically. Be extra
vigilant when using continuous mode on neonates or
hemodynamically compromised patients.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...