A
LARM
S
ETUP
(A
LARM
L
IMITS
T
ABLE
)
VF8
D
ELTA
/D
ELTA
XL/K
APPA
5-9
et ENF/i ENF
0 to 7.5 %
Off
Low 0% High 6%
et HAL/i HAL
0 to 7.5 %
Off
Low 0% High 6%
et ISO/i ISO
0 to 7.5 %
Off
Low 0% High 6%
et O
2
10 to 100 %
Off
Low 10% High 100%
et SEV/i SEV
0 to 9.0 %
Off
Low 0% High 9%
etCO
2
/etCO
2
*
5 to 95 mmHg (0.7 to 12.6 kPa)
Off
Low 30 mmHg (4.0 kPa)
High 50 mmHg (6.7 kPa)
FiO
2
18 to 100 %
On
Low 18% High 100%
GP1/GP2
S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
S: (Adult/Pediatric)
Low 90 mmHg (12.0 kPa)
High 160 mmHg (21.3 kPa)
M: (Adult/Pediatric)
Low 60 mmHg (8.0 kPa)
High 125 mmHg (16.7 kPa)
D: (Adult/Pediatric)
Low 50 mmHg (6.7 kPa)
High 110 mmHg (14.7 kPa)
S: (Neonatal)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Neonatal)
Low 40 mmHg (5.3 kPa)
High 85 mmHg (11.3 kPa)
D: (Neonatal)
Low 35 mmHg (4.7 kPa)
High 80 mmHg (10.7 kPa)
HR
20 to 300 beats per minute
On
Adult:
Low 45
bpm
High
120 bpm
Pediatric:
Low 50
bpm
High
150 bpm
Neonatal:
Low 80
bpm
High
170 bpm
iCO
2
/iCO
2
*
2 to 10 mmHg (0.3 to 1.3 kPa)
(upper high limit only)
Off
High: 4 mmHg (0.5 kPa)
ICP/ICP2/
ICP3/ICP4
-25 to 300 mmHg (-3.3 to 40.0 kPa) Off
Low 2 mmHg (0.3 kPa)
High 20 mmHg (2.7 kPa)
iO
2
18 to 100 %
Off
Low 18% High 100%
LA
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
Low 0 mmHg (0.0 kPa)
High 20 mmHg (2.7 kPa)
Parameter
Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...