Cleaning the Sidestream Pump
To clean the sidestream pump
1. Set etCO
2
measurement mode to
Side
(for Sidestream monitoring).
2. Remove the etCO
2
module or pod from the monitor.
3. Remove all sidestream sampling tubing from the module or pod connectors.
4. Attach the section of 1/8- or 3/16-inch tubing to the exhaust (sidestream
output) port on the module or pod, and run it to a drainage receptacle placed
below the module.
5. Fill the 60cc catheter-tip syringe with cleaning fluid, and fix it to the
sidestream input connector on the etCO
2
module or pod.
6. Flush the fluid slowly through the pumping system and out through the tubing
connected to the exhaust port. Repeat two more times for a total of 180cc of
fluid.
7. Remove the syringe. Leave remaining fluid in the pumping system for 30
minutes. This disinfects the system.
8. After 30 minutes, fill the syringe with distilled water and flush through the
system. Repeat two more times.
9. Empty the syringe and use it to push several volumes of air slowly through the
system. This clears most of the solution from the pump.
10. Repeat step 9 one or more times to make sure that as much fluid as possible
has been cleared from the system.
11. Remove the syringe from the module or pod, but keep the drain tubing in
place.
CAUTION:
Always use a syringe to flush cleaning solutions through the pump as described
in the instructions below.
Do not attempt to use the sidestream sampling pump itself to move cleaning
solutions through the system. This may cause accelerated wear on the pump
bearings.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...