Tubing Connections
To connect the module and tubing
1. Connect one end of the sampling tubing to the water trap, and the other end to
the airway T-connector. For information on changing sampling tubing and T-
connector, see page 23-24.
2. If return of sample gas is not possible, connect one end of the exhaust tubing
to the exhaust port at the rear of the module (see illustration on page 23-5),
and the other end to the hospital’s gas-scavenging system.
3. Connect recirculating sample gas tubing as follows:
For Dräger COSY breathing system (Fabius GS):
- Use the sample gas return kit.
- Push the rubber sleeve on to the exhaust port on the back of Scio and
plug the connector into the socket on the front of the COSY (Fabius
GS) until it clicks into place.
For other breathing systems:
-Use the sample gas return kit with an integrated bacterial filter and
connect it to the breathing system used adjacent to the expired gas
valve so that the returned sample gas is routed through the CO
2
absorber.
NOTE:
Do not use the module without a water trap.
NOTE:
If possible, use sample gas recirculation for the following reasons:
To prevent increased anesthetic agent concentration in the operation room.
To conserve anesthetic agents.
To prevent undesired losses of volume during low-flow application.
WARNING: Accumulation of exhaled metabolic products may
occur during low flow anesthesia, therefore monitoring of
oxygen concentration is mandatory. Flushing the circuit with
fresh gas at regular intervals may be necessary particularly if
decreasing oxygen concentrations are measured.
CAUTION:
Replace the bacterial filter in the sample gas tubing every six months.
Strictly follow Instructions for Use of the breathing system.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...