WARNING:
Occupational safety: Used sampling tubing, T-
connectors, and water traps could be contaminated and
must be handled and disposed of with care. Infection
hazard may be present. Dispose of these items in
accordance with local regulations.
Optimize response time by reducing dead space and by
keeping sampling tubing as short as possible without
stretching it. Long sampling lines may decrease
accuracy and cause slower response times for side-
stream measurement technologies.
When using a sampling line for intubated patients with
a closed suction system, do not place the airway
adapter between the suction catheter and endotracheal
tube. This is to make sure that the airway adapter does
not interfere with the functioning of the suction
catheter.
CAUTION:
Leaks in the breathing circuit (for example, an uncuffed endotracheal tube or a
damaged airway adapter) may significantly affect CO
2
and respiratory
mechanics measurement values (for example, airway pressure, flow, dead
space, and CO
2
production).
Calibrate the sensor when you switch sensor types (for example, from adult to
neonatal).
To prevent damage to sensor windows, remove the airway adapter from the
circuit whenever aerosolized medication is being delivered.
To prevent leakage, make sure the airway adapter is firmly connected to the
breathing circuit.
Cutting or removing any part of the sample line could compromise
measurement accuracy.
In high-altitude environments, etCO
2
values may be lower than values observed
at sea level, as described by Dalton's law of partial pressures. When using the
monitor in high-altitude environments, it is advisable to adjust etCO
2
alarm
settings accordingly.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...