S
TATUS
M
ESSAGES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
17-19
SpO
2
Searching
The monitor has not been able to
compute a valid measurement and is
searching for a pulse.
• Make sure the SpO
2
sensor is
properly attached to the patient at a
site free of patient motion and all
cables are properly connected.
• Make sure the correct sensor type
and/or site is used.
• Contact DrägerService.
SpO
2
Unplugged
The sensor cable not connected to
the MultiMed pod.
• Make sure the cables are securely
connected.
• Check for a defective sensor.
SpO
2
: Pod Failure
Pod hardware or software failure.
• Disconnect/reconnect the pod if
problem persists.
• Disconnect/reconnect the pod and/
or replace with a good sensor.
• Contact the hospital’s technical
personnel (SpO
2
compatibility is a
locked option).
SpO
2
: Check/
replace Sensor
Defective/faulty sensor
• Replace with a Nellcor-compatible
sensor.
• Contact the hospital’s technical
personnel (SpO
2
compatibility is a
locked option).
SPO
2
: non-Nellcor
compatible sensor
A non-Nellcor compatible sensor or
SpO
2
pod is connected to a monitor
that is configured for Nellcor.
• Replace with a Nellcor-compatible
sensor.
• Contact the hospital’s technical
personnel (SpO
2
compatibility is a
locked option).
Duplicate Device
Connected
A MultiMed pod (with SpO
2
sensor)
and an Infinity Nellcor Oximax pod
are connected simultaneously.
Disconnect the duplicate device.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...