![Dräger Infinity Delta Instructions For Use Manual Download Page 443](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779443.webp)
O
VERVIEW
VF8
D
ELTA
/D
ELTA
XL/K
APPA
26-5
Body Surface Area (BSA), Predicted Body Weight (PBW),
Predicted Body Surface Area (PBSA)
The input for height, weight, patient category, and sex are needed for the calculation
of the body surface area (BSA), predicted body weight (PBW) and predicted body
surface area (PBSA). These calculated body characteristics are required for the
indexing
of PiCCO parameters. For the indexing of volumetric parameters, the PiCCO
pod uses the following calculated parameters which are dependent on the following
body characteristics:
BSA = Body Surface Area (m
2
)
PBW = Predicted Body Weight (kg); calculated according to height, sex, and
patient category
PBSA = Predicted Body Surface Area (m
2
); calculated with the PBW instead
of actual body weight.
Calculation
Category
Sex
PBW (kg) = 50 + 0.91 [height (cm) – 152.4]
Adult
(habitus)
1
Male
PBW (kg) = 45.5 + 0.91 [height (cm) – 152.4]
Female
PBW (kg) = 39 + 0.89 [height (cm) – 152.4]
Pediatric (>152.4 cm)
(habitus)
1
Male
PBW (kg) = 42.2 + 0.89 [height (cm) – 152.4]
Female
PBW (kg) = ((height (cm))
2
x 1.65) / 1000
Pediatric (<152.4 cm)
Male and female
1
NOTE
: In marginal cases the decision about the correct patient category for the individual patient
must be based on the anatomy and appearance (habitus).
NOTE:
PBW is used to calculate the parameter EVLWI. PBSA is used to calculate
GEDVI/ITBVI while BSA is used to calculate all other index parameters.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...