TP Cal Required
The sensor must be calibrated before
use and after the following:
• Change in sensor temperature setting
• Change in site timer setting
• Change in manually entered
barometric pressure value
• Change in auto heater shutdown
setting
Calibrate the sensor.
TP Cal Accepted
The sensor calibration was successful. Apply the sensor and proceed with
monitoring.
TP Cal Failed
The sensor calibration was
unsuccessful.
Recalibrate the sensor.
TP Barometric
Cal Failed
The pod’s internal barometric pressure
sensor needs calibration.
Contact DrägerService.
TP Comm
Failure
Loss of communication between
monitor and pod.
• Remove the sensor from the patient
immediately.
• Check the pod and the cable and
connections. If necessary, replace
the cable or pod.
TP Drift
The actual sensor temperature differs
from the target temperature setting by
> 0.3 °C.
• Make sure the sensor is securely
placed in the fixation ring and ring is
properly attached to the patient.
• Replace the sensor if necessary.
TP Cartridge
(pod)
Incompatible
The tpO
2
/CO
2
pod is incompatible with
the monitor.
• Replace the pod.
• Contact DrägerService.
TP Pods Only
Supported on
AUX1 or AUX2.
TP pod is connected to the wrong
connector at the monitor.
Connect the TP pod to one of the
bottom two pod connectors of the
monitor (Hemo/Aux. or Aux.).
TP Pod H/W
Failure
The tpO
2
/CO
2
hardware failure.
• Remove the sensor immediately
from the patient.
• Check the sensor; the pod and the
pod connections. Replace the sensor
or the pod, if necessary.
• Disconnect then reconnect the pod.
• Recalibrate the sensor.
• If problem persists, call
DrägerService.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...