Overview
The monitor can acquire and process non-invasive blood pressure (NBP) signals and
display the results. Blood pressure measurements are determined by the oscillometric
method and are equivalent to those obtained by intra-arterial methods, within the
limits prescribed by the Association for Advancement of Medical Instrumentation,
Electronic Automated Sphygmomanometers (AAMI/ANSI SP-10).
The monitor’s NBP system inflates and then deflates a pneumatic cuff wrapped
around the patient’s arm or leg. A hose links the cuff to the monitor, which determines
systolic, diastolic, and mean blood pressures for adult, pediatric, or neonatal patients.
The monitor can initiate blood pressure measurements singly, at set intervals, or
continuously over a 5-minute period.
The monitor can be configured to sound an attention tone whenever an NBP
measurement is completed (see page 2-9 for information).
Display Features
The monitor displays non-invasive blood pressure in the form of numerical values and
trends. There is no NBP waveform. For information on trended data, see chapter 6,
Trends.
The NBP parameter box shows the latest readings for mean, systolic and diastolic
pressure. See page 2-5 for more information on prioritizing and displaying parameter
boxes.
NOTE:
NBP should be calibrated yearly by your hospital’s technical personnel or
other qualified personnel as described in the monitor’s Service manual (see “Safety
Considerations” on page 6).
1
Systolic pressure
2
Alarm limits (crossed bell icon
if alarm is disabled)
3
Diastolic pressure
4
Mean pressure
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...