WARNING:
The NeoMed and MultiMed 12 pods are not intended for
use during electrosurgery. To protect patients from
burns, do not use these pods in an ESU environment.
For better performance and to reduce the hazard of
burns during surgery, always use accessories designed
for ESU environments.
To reduce the hazard of burns during surgery, keep the
sensor or transducer (ECG, temperature, pressure,
SpO
2
, BISx) and their associated cables away from the
surgical site, the electro-surgical unit return electrode,
and earth ground.
Always use a Dräger ESU block or MultiMed Plus OR
cable with compatible lead wires. Doing so reduces
ESU interference and protects the patient from burns
caused by ESU-induced current flowing through the
lead wires. For better performance, also set the ECG fil-
ter option to ESU.
Dräger recommends using the ESU block during
electrosurgery. If you do not have an ESU block or a
MultiMed Plus OR, use only Dräger blue ECG lead sets.
They help protect the patient from burns caused by
ESU-induced current flowing through the leads.
While the ESU block or the MultiMed Plus OR cable are
in use, impedance respiration monitoring is inoperative
and the detection of pacemaker spikes is degraded. If
pacemaker detection is enabled, the ESU interference
may be detected as pacemaker spike.
Dräger recommends the use of the MultiMed Plus OR
during electrosurgery only.
Do not use the MultiMed Plus OR cable with Dräger blue
ECG lead wires. Doing so, will degrade performance
which can result in inaccurate values.
NOTE:
Use SpO
2
instead of the ECG parameter to determine heart rate.
Use rectal temperature probe sheaths to cover internally placed temperature
sensors.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...