NBP S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
13-3
NBP Setup
Safety Considerations
Software and Hardware Cuff Pressure Cutoffs
The cuff deflates automatically if a measurement takes longer than 2 minutes in
Adult/
Pediatric
mode or 90 seconds in
Neonatal
mode. To protect the patient from
excessive pressure, inflation limits have been established in all patient categories; see
Appendix B, for more information.
Cuff Selection and Placement
The quality of NBP monitoring depends largely on the quality of the signals received
by the monitor. It is important to select the correct cuff size for your patient. Measure
the circumference of your patient’s limb and use the markings on the cuff to select the
correct cuff size.
WARNING: Obstructions may cause the cuff to inflate and
deflate improperly and result in inaccurate readings. Check
the hose and cuff for damage and dirt. Do not allow the hose
and cuff to come in contact with fluids, and make sure that
they are not compressed or kinked.
WARNING: Accurate NBP meas urements depend on the
correct size and type of the blood pressure cuff in relation to
the patient's arm circumference. The wrong sized cuff, or
cuffs outside the range or size manufactured by Dräger, can
cause inaccurate measurements. Use only Dräger approved
cuffs and make sure that the correct size is used for each
patient.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...