24-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
24 N
EUROMUSCULAR
T
RANSMISSION
(NMT)
MONITORING
NMT Check sensor
The sensor is malfunctioning
or disconnected.
• Check the sensor application.
• Replace the acelerometer sensor, if
necessary.
NMT Artifact
• Excessive signal noise due
to patient’s movement/
tremor
• Interference from auxiliary
equipment
• Check the electrodes and reapply, if
necessary.
• Isolate the patient from auxiliary
equipment, if necessary.
• Isolate the patient from extraneous
movement or vibration.
NMT Check temp
sensor
The temperature cable is
broken or missing.
Check the sensor and replace, if
necessary.
NMT Temp Out of
Range (High/Low)
The temperature value is
outside the measuring range
(high/low).
• Check the patient and treat, if
necessary.
• Check the sensor and replace, if
necessary.
NMT
TOF-Cnt > UL
TOF-Cnt < LL
The number of twitches fall
outside the set alarm limits.
• Check the patient and treat, if
necessary.
• Change the alarm limits as
appropriate.
NMT Single Out of
Range (High)
The single mode parameter
value is outside the measuring
range (high).
Check the patient and treat, if
necessary.
NMT Pod H/W Failure
• Trident pod hardware failure
• An unsupported pod is
connected.
• Check the NMT pod connection.
• Disconnect the unsupported pod and
reconnect the NMT pod.
• Call the hospital’s technical personnel
or DrägerService.
NMT Pod
disconnected
The NMT pod is not
connected to the monitor.
• Check the cables and connections.
• Replace the cables, if necessary.
Duplicate device
connected
More than one NMT pod is
connected.
Remove the secondary pod.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...