12-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
12 R
ESPIRATION
To display the OCRG screen
1. Set the patient type to
Neonatal
(see page 2-4).
2. Connect SpO
2
sensors, HR leads, and respiration or etCO
2
leads.
3. Set the apnea time on the RESP menu (see page 12-7).
4. Press the
Fast Access
fixed key.
5. Click on
OxyCRG
to display the OCRG screen.
Scale
To change the bbHR scale
1. Using the rotary knob, highlight the value at the top left corner of the grid and
click.
2. Dial to desired scale setting and click.
Values are shown in the following table (you can modify only the HR scale).
Cursor
When you click on
Cursor
, a vertical bar appears on the trend area of the screen, and
cursor time is displayed in the upper right corner. The numbers on the left side of the
screen no longer represent scale values, but rather parameter values at the time marked
by the cursor. The monitor continues to display current (real-time) values on the right
of the screen. When you move the cursor to the right or the left with the rotary knob,
the cursor-time values and the cursor time are modified and displayed accordingly.
Parameter Scale
Definition
bbHR
10-180 bpm (lowest setting)
130 - 300 bpm (highest setting)
Highest (max) and lowest (min) bbHR value over
the last three minutes
SpO
2
50 -100 %
Lowest saturation value over last three minutes
tpO
2
/
tpO
2
*
30 - 110 mmHg
(4 - 16 kPa)
Lowest tpO
2
/tpO
2
* saturation value over last three
minutes
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...