![Dräger Infinity Delta Instructions For Use Manual Download Page 423](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779423.webp)
M
ONITORING
M
ODES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
24-5
Monitoring Modes
Single Twitch Mode
In this mode, the pod sends a single stimulation pulse and
measures the magnitude of the resulting muscle twitch. The
pod then calculates an NMT value (from 0 to 100 %) by
comparing this magnitude to that of the previously
established reference twitch (see page 24-2). For example:
If the displayed value is 79, the magnitude of the measured
twitch was 79 % of that of the reference twitch.
A bar graph in the NMT parameter box indicates the
relative magnitude of the twitch. The pod repeats the
stimulation automatically at regular, menu-selectable
intervals (default is 20 seconds).
If the pod is unable to establish a reference twitch, it uses an
internal reference value for single-mode measurements.
Train-of-Four (TOF)
In this mode (default), the pod sends a sequence of four
stimulation pulses 0.5 seconds apart and measures the
magnitude of the resulting muscle twitch after each
individual pulse. The pod then calculates a TOF response
value (TOF-ratio, from 4 to 100 %) using the ratio between
the fourth twitch and the first twitch. A bar graph in the
NMT parameter box indicates the relative magnitude of each twitch. The pod repeats
this stimulation sequence automatically at regular, menu-selectable intervals (default
is 20 seconds).
When fewer than four twitches are detected, the monitor
displays a TOF-Count (TOF-Cnt) value instead, which
represents the number of responses to the four TOF
stimulation pulses.
CAUTION:
NMT stimulation can be painful to a non-sedated patient. Do not stimulate
before adequate sedation.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...