![Dräger Infinity Delta Instructions For Use Manual Download Page 142](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779142.webp)
8-2
D
ELTA
/D
ELTA
XL/K
APPA
VF8
8 ECG
AND
H
EART
R
ATE
Overview
The monitor can calculate heart rate, detect arrhythmia conditions (adult and pediatric
patients only), and display ECG data. Lead wires are connected to the monitor via
special pods designed to facilitate cable management. The MultiMedCables
accommodate standard (3-lead, 5-lead, or 6-lead) cable sets with the (MultiMed 12
Pod) or a 12-lead cable set with the (optional MultiMed 12 Pod). The NeoMed
Pod
is
designed for 3-lead monitoring for neonatal patients, allowing one ECG waveform to
be displayed. The MultiMed and NeoMed cables also have connectors for an SpO
2
sensor and up to two temperature probes. A connector for an FiO
2
sensor is available
only on the NeoMed cable.
The MultiMed cable also has connectors for an SpO
2
sensor and up to two
temperature probes. The NeoMed pod has all these connectors plus a connector for an
FiO2 sensor. The SpO
2
port is not enabled if you have a Masimo SET or Nellcor
OxiMax logo on the front of your monitor.
Before you begin ECG and heart rate monitoring, proceed as follows:
1. Connect the appropriate MultiMed or NeoMed cable to the monitor.
2. Plug leads and accessories into designated connectors.
3. Attach the lead wires to the electrodes on the patient.
1
MultiMed 5 cable with accessories (NeoMed, MultiMed6, MultiMed Plus,
and MultiMed Plus OR cables look similar)
2
MultiMed 12 pod (Connects to the PodCom connector (Not in scale to
MultiMed 5 pod)
Delta/Delta XL
Kappa
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...