![Dräger Infinity Delta Instructions For Use Manual Download Page 474](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779474.webp)
27-6
D
ELTA
/D
ELTA
XL/K
APPA
VF8
27 B
ODY
T
EMPERATURE
Temperature Setup
To access the temperature setup menu
Click on the desired temperature parameter box (if displayed);
or
1. Press the
Menu
fixed key.
2. Click on
Patient Setup
.
3. Click on
Parameters
.
4. Click on
TEMP
,
TEMP2
or
TEMP3
,
depending on which temperature signal
you wish to monitor (see page
The TEMP setup menu displays only two items:
TEMP display
– Configures the bottom half of the parameter box to display
either the reading of the second temperature probe (b) or the difference
between the first probe’s reading and the second (
DT,
the delta value).
TEMP alarms
– Accesses temperature alarm settings on the Alarm Limits table
Status Messages
Message
Possible Cause
Suggested Action
> #
< #
The temperature exceeds the
upper or lower alarm limits.
• Check the patient and treat, if
necessary.
• Check equipment and replace if
necessary.
Out of Range (High)
Out of Range (Low)
The temperature value is greater
or less than the measuring range.
Can’t Derive
Δ
T(#)
The cable is defective or
unplugged.
• Check equipment and replace if
necessary.
• Connect the second temperature
probe.
Unplugged
Check the equipment and replace if
necessary.
H/W Failure
Temperature circuitry failure.
Contact DrägerService.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...