6. Press the Start/Stop key (
) on the calibration unit to start the flow of gas,
checking the gauge reading on the front of the unit to verify an adequate gas
supply.
7. Access the tpO
2
/CO
2
setup menu.
8. Click on
Start Calibration
.
or
— Press and hold the calibration key on the front of the tpO
2
/CO
2
pod for
two seconds.
— Calibration information appears on the right side of the menu. The moni-
tor sounds a tone when calibration is complete.
9. Remove the sensor from the calibration chamber, holding it away from
sources of CO
2
. The monitor performs a final response test of the CO
2
sensor.
If the calibration is successful, the date and time of the calibration are displayed on the
tpO
2
/CO
2
menu, and the site timer, once it is enabled, starts counting down the timed
monitoring session.
Calibration typically takes 3 to 5 minutes. The monitor performs the following
functional checks:
— Temperature Test — Turns on the sensor heater and verifies the sensor heats
to the set temperature (±0.05 °C) within 3 minutes
— Range Test — Verifies the sensor produces voltages within acceptable ranges
when exposed to the reference gas
— Stability Test — Verifies the sensor produces stable voltages for at least 1
minute after exposure to the reference gas
NOTE:
The calibration unit automatically cuts off the flow of reference gas after 20
minutes, but the gas flow can be stopped before that by pressing the
Start/Stop
key.
NOTE:
If you do not remove the sensor within 30 minutes after the start of
calibration, the sensor heater switches off and a new calibration is required.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...