![Dräger Infinity Delta Instructions For Use Manual Download Page 175](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779175.webp)
S
TATUS
M
ESSAGES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
9-9
Status Messages
Archive
Determines whether the selected event is
stored, recorded, or both. You can view
stored events on the Event Recall screen
(see page 1-27).
NOTE
: The archive function for ASY
(Asystole) and VF (Ventricular Fibrillation)
cannot be disabled.
•
Store
; stores the selected event
•
Record
; generates an alarm recording
of selected event
•
Str./Rec.;
stores an event and
generates an alarm recording (even
when event alarm is turned off)
• OFF
Message
Event Definition
Suggested Action
PVC/min > UL
The PVC/min value is above the upper alarm
limit.
Check the patient.
Cannot Learn
Lead <X>
At the end of the Learning phase, the
dominant normal complex could not be
determined for Lead <X> (one of the two
leads selected for QRS processing).
Arrhythmia analysis proceeds using the other
lead as a source. Lead <X> is ignored until a
Relearn
is initiated.
• Check the leads.
• Choose another lead for QRS
processing.
Relearning
Monitor is learning a normal QRS complex as
a reference template.
Not applicable
Unable to
Learn
After 100 beats, monitor cannot determine
the dominant normal complex on any lead
selected for QRS processing. Learning
continues.
Check the electrode preparation.
Baseline
Artifact
Artifact is blocking arrhythmia classification.
Check the electrode preparation.
Function
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...