Frank Lead Format
The Frank lead format shows cardiac activity in three mutually
perpendicular directions. Seven Frank electrodes (I, E, C, A, M,
H, and F) are combined to produce Frank leads X, Y, and Z (the F
electrode is not necessary for Frank leads). Using this format, a
cardiologist can work with leads X, Y, and Z at the CRS
(Cardiology ReviewStation). However, the bedside and the central
station use the standard and more familiar 12-lead format.
The following tables show Frank electrodes used to derive Frank
leads X, Y, and Z, and the correlation between Frank electrode
positions and chest electrode positions.
When you select
Frank
monitoring (see “Twelve-Lead Configurations (Standard and
The displayed limb leads are true limb leads. The chest leads remain derived.
Frank leads are converted to standard 12-lead format.
If a Frank electrode falls off, the monitor displays a
Lead Off
message and the
corresponding Frank electrode label.
Devices connected to the monitor via the network (central station, printers,
CRS) work with data in the 12-lead format.
The CRS can display either the 12-lead or Frank format.
Rest-ECG reports are labeled
Derived 12-lead
.
Frank electrodes used
to derive Frank leads
Correlation between chest and Frank
electrode positions
Frank Electrode
Frank Lead Label
Chest Electrode
Frank Position
A, C, I
X
V1
I
F, M, H
Y
V2
E
A, M, I, E, C
Z
V3
C
V4
A
V5
M
V6
H
NOTE:
Only leads I and II are true limb leads (lead III is calculated). If the RA
electrode is disconnected, lead III appears as a flat line to alert the clinician to the
disconnected electrode. All other leads function as expected.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...