![Dräger Infinity Delta Instructions For Use Manual Download Page 576](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779576.webp)
B-50
D
ELTA
/D
ELTA
XL/K
APPA
VF8
B T
ECHNICAL
D
ATA
etCO
2
/Respiratory Mechanics
Pressure Parameters
Measurement range,
Adult
PIP
PEEP
tPEEP
MAP
Pause
0 to 120 cmH2O
0 to 98 cmH2O
0 to 98 cmH2O
0 to 120 cmH2O
0 to 120 cmH2O
Measurement range,
Pediatric
PIP
PEEP
tPEEP
MAP
Pause
0 to 120 cmH2O
0 to 98 cmH2O
0 to 98 cmH2O
0 to 120 cmH2O
0 to 120 cmH2O
Measurement range,
Neonatal
PIP
PEEP
tPEEP
MAP
Pause
0 to 120 cmH2O
0 to 98 cmH2O
0 to 98 cmH2O
0 to 120 cmH2O
0 to 120 cmH2O
Flow/volume parameters
Measurement range,
Adult
PIF
PEF
TVi
TVe
RRv
MVe
TV lk
0 to 180 L/min
0 to 180 L/min
0, 100 to 3000 mL
0, 100 to 3000 mL
0 to 150 1/min
0 to 60 L/min
0 to 100 %
Measurement range,
Pediatric
PIF
PEF
TVi
TVe
RRv
MVe
TV lk
0 to 120 L/min
0 to 120 L/min
0, 10 to 3000 mL
0, 10 to 3000 mL
0 to 150 1/min
0 to 60 L/min
0 to 100 %
Measurement range,
Neonatal
PIF
PEF
TVi
TVe
RRv
MVe
TV lk
0 to 28 L/min (etCO2/flow),
0 to 40 (flow only) (PIF& PEF)
0, 100 to 500 mL
0, 100 to 500 mL
0 to 150 1/min
0 to 10 L/min
0 to 100 %
Measurement range,
Adult/Pediatric
TVi s, TVi m, TVe s, TVe m
MVe s, MVe m
RRs, RRm
0, 100 to 3000 mL
0 to 60 L/min
0 to150 1/min
Measurement range,
Neonatal
TVi s, TVi m, TVe s, TVe m
MVe s, MVe m
RRs, RRm
0, 2 to 500 mL
0 to 10 L/min
0 to 150 1/min
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...