IBP Setup
IBP setup is a two-phase process involving the following procedures. After
configuring individual IBP parameters, you must assign them to connected IBP
channels.
To access an IBP parameter setup menu
Click on the respective parameter box on the main screen;
or
1. Press the
Menu
fixed key to display the Main menu.
2. Click on
Patient Setup
.
3. Click on
Parameters
.
4. Scroll to the IBP parameter you want to configure (ART, LV, PA, RV, CVP,
RA, LA, ICP, ICP2, ICP3, ICP4, GP1, or GP2) and click. The setup menu
appears, with the name of the parameter you have selected displayed at the top
of the menu.
Quick Reference – IBP Setup
Available functions, present on all IBP setup menus, are described in the following
table:
IBP Setup Menus
Menu Item
Description
Available Settings
Zero
Zeroes the transducer and displays
time and date of last zeroing operation
(see page 14-3).
Not applicable (read-only)
Scale
Sets the upper values of the IBP
waveform scale.
• ART, CVP, LV, GP1, GP2, ICP,
ICP2, ICP3, ICP4, LA,
P1-3 (a-d), PA, RA and RV: 5,
10, 15, 20, 25, 30, 35, 40, 50,
75, 100, 125, 150, 175, 200,
225, 250,
and
300 mmHg
Filter
Adjusts the filter applied to the IBP
signal.
8, 16,
and
32 Hz
Cal Factor
Determines the calibration factor.
80 – 120
Last Cal Factor
Displays the time of the last successful
calibration.
Not applicable (read-only)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...