Status Messages
EEG Mini
Trend
Allows you to select parameters for
display in mini-trends (see page 6-7
for more information on mini-trend
display)
1) Press the
Fast Access
fixed key.
2) Click on
Split Screen
.
3) Click on
10 Min Trends
or
60 Min
Trends
.
4) Open the EEG menu as described on
page 11-6.
5) Click on the setting of your choice:
Med, SEF, BSR, Power, Beta, Alpha,
Theta,
or
Delta
SEF Setting
Determines Spectral Edge Frequency
(SEF
)
level
• 75, 85, 90, 95, 97, or 98 %
Low Filter
1, 2
Allows user to compensate for
excessive baseline drift.
•
0.5 Hz
(Default)
• 1.0 Hz
High Filter
1, 2
Allows user to compensate for
electrical noise artifact.
• 15 Hz
• 30 Hz
(Default)
• Off
Notch Filter
1, 2
Allows user to compensate for power
line noise artifact.
• On
(Default)
• Off
1
Caution: Only qualified personnel should access Filter functions.
2
Selection is grayed out if EEG pod is software version VE0.
Message
Possible Cause
Suggested Action
EEG Check
Electrodes
An EEG channel has
degraded, suggesting poor
electrode contact.
Replace the electrodes.
EEG Pod
Initializing
The EEG pod is warming up.
Wait until the pod has warmed up before
attempting to measure.
EEG Pod
Disconnected
The EEG pod is not connected
to the monitor.
Check the cable and connection. Replace the
cable, if necessary.
EEG Pod
Software Failure
An EEG pod software problem
has been detected.
Call your hospital’s technical personnel or
DrägerService.
EEG Pod
Hardware Failure
Loss of communication or
hardware problem detected.
• Check that the EEG pod software is
compatible with the monitor software.
• Check the EEG pod connection.
• Call your hospital’s technical personnel or
DrägerService.
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...