. The Warmer2 menu has the following selections.
Babytherm Status Messages
Menu Item
Description
Available Settings
Show All
Parameters
Displays all parameters
# Parameter
Boxes
Select number of Parameter Boxes
1
,
2
(default)
NOTE
: Warmer2 Parameter box is
not displayed when the number of
Parameter Boxes is set to 1
Parameter 1
Selects up to 3 parameters for display in
warmer parameter box
T1skin
,
T2skin
,
Tmatt
,
HtrPwr
(default),
None
Parameter 2
T1skin
,
T2skin
,
Tmatt
,
HtrPwr
,
None
(default)
Parameter 3
T1skin
,
T2skin
,
Tmatt
,
HtrPwr
,
None
(default)
Warmer
Mini Trend
Allows you to select parameters for
display in mini trends (See page 6-7 for
more information on mini trend display).
T1skin
,
T2skin
,
Tmatt
,
HtrPwr
(default)
Message
Condition
Suggested Action
Duplicate Device
Connected
Multiple devices are connected.
Make sure only one incubator is
connected to the monitor.
T1skin Out of Range
The T1skin value is out of range
(high or low).
• Check the patient and treat, if
necessary.
• Check the connections.
• Unplug and reconnect the
incubator.
• Power-cycle the monitor or
undock and redock the monitor.
• Call the manufacturer.
T2skin Out of Range
The T2skin value is out of range
(high or low).
Tmatt Out of Range
The Tmatt value is out of range (high
or low).
HtrPwr Out of Range
The HtrPwr value is out of range
(high).
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...