Refer to the “Safety Considerations” section starting on page 6, at the beginning of
these Instructions for Use, for general precautions regarding safe device operation.
WARNING:
The NMT pod's high frequency pulses may interfere
with other sensitive equipment such as cardiac
pacemakers. Do not use the NMT pod on patients with
implanted medical devices unless so directed by a
medical specialist.
To avoid adversely affecting the pod, do not use the
pod in close proximity to shortwave or microwave
therapy devices.
Although the NMT pod is designed for use with
electrosurgical equipment, in unusual circumstances
simultaneous use of an NMT pod with such equipment
may result in burns at the stimulation site and damage
to the NMT pod.
Use peripheral nerve stimulator electrodes with
circular conductive areas between 1 cm (.39 in) and
1.5 cm (.59 in) in diameter. Apply these electrodes
according to the Instructions for Use. Pay special
attention to current densities > 2 mA r.m.s./cm
2
.
Stimulation voltage should not exceed the maximum
specified by the electrode manufacturer.
To avoid serious injury to the patient, apply the
stimulation electrodes close together as described. Do
not apply the electrodes trans-thoracically (across the
chest) or transcerebrally (across the head). Doing so
can cause electrical current to enter the chest or head
and may lead to irregularities in cardiac rhythm, brain
activity, or pain.
CAUTION:
The NMT pod’s stimulus indicator flashes when stimulation is occurring. To
avoid being shocked unintentionally, do not touch the stimulation electrodes
when it is flashing.
Do not use NMT skin temperature sensors to monitor core body temperature.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...