S
TATUS
M
ESSAGES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
25-9
— Replace the sensor, if it continues to fail.
*** -
if an electrode is not making contact (off) or if an electrode's impedance
value is high or clipping. In this case, press on the electrode to improve
contact and make sure its impedance value is within limits.
Status Messages
NOTE:
The fourth electrode's impedance value and status are blank if a sensor with
three electrodes (for example, a pediatric sensor) is used.
Message
Possible Cause
Suggested Action
BIS: Pod Initializing
The BISx pod is initializing
and performing a self-test.
Wait until the initialization is completed.
BIS: Sensor
impedance check in
progress
The BISx pod is performing a
sensor impedance check.
• Wait until the impedance check is
completed.
• Cancel and retry.
BIS: Sensor check
failed
• The BISx pod failed the
sensor impedance check.
• Defective sensor cable.
• Retry the impedance check.
• Re-prep the electrodes and retry the
impedance check.
• Replace the sensor, if necessary, and
perform a manual impedance check.
• Replace the sensor cable.
BIS > UL
BIS < LL
The BIS parameter value is
above higher/below lower
alarm limit.
• Check the patient and treat, if
necessary.
• Change the alarm limits as
appropriate.
BIS: EMG Out of
Range (High)
The EMG parameter value is
out of range high.
• Check the sensor.
• Isolate the patient from auxiliary
equipment, if necessary.
• Isolate the patient from extraneous
movement or vibration.
• Check the patient and treat, if
necessary.
BIS: PWR Out of
Range (High/Low)
The PWR parameter value is
out of range high/low.
BIS: BCT Out of
Range (High)
BCT parameter value out of
range high/low.
BIS: Signal quality
poor, SQI < 50%
• Artifact caused by motion or
eye blinks.
• Interference from electro-
surgical unit.
• Electrode loosened from
patient.
• Check the sensor.
• Isolate the patient from extraneous
movement or vibration.
• Isolate the patient from auxiliary
equipment, if necessary.
• Re-prep the patient.
BIS: SQI too low,
(SQI value < 15%)
BIS: Sensor
unplugged
• Sensor malfunctioning or
disconnected.
• Defective sensor cable.
• Check the cables and connections.
• Replace the sensor, if necessary.
• Replace the sensor cable.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...