![Dräger Infinity Delta Instructions For Use Manual Download Page 442](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779442.webp)
26-4
D
ELTA
/D
ELTA
XL/K
APPA
VF8
26 P
ULSE
C
ONTOUR
C
ARDIAC
O
UTPUT
(P
I
CCO)
MONITORING
The transpulmonary thermodilution PiCCO parameters are listed in the following
table.
Transpulmonary Thermodilution parameters
Label
Parameter
Measurement
Range
Normal Value
p-CO
Cardiac Output - Transpulmonary
thermodilution cardiac output serves as the
basic parameter for calculation of various
blood volumes.
0.25 to 25 L/min
Not applicable
p-CI
Cardiac Index - Cardiac output normalized to
BSA (Body Surface Area).
0.10 to
15.00 L/min/m
2
3.0 - 5.0 L/min/m
2
GEDV
Global End Diastolic Volume - The sum of all
end-diastolic volumes of the atria and the
ventricles. GEDV is equivalent to preload
volume of the total heart.
40 to 4800 ml
Not applicable
GEDVI
Global End Diastolic Volume Index
GEDVI = GEDV / PBSA (Predicted Body
Surface Area)
80 to 2400 ml/m
2
680 to 800 ml/m
2
EVLW
Extravascular Lung Water - correlates to
extravascular thermal volume in the lungs and
is determined by the mean transit time
method.
10 to 5000 ml
Not applicable
EVLWI
Extravascular Lung Water Index -
Extravascular Lung Water normalized to body
weight.
EVLWI = EVLW / PBW (Predicted Body
Weight)
0 to 50 ml/kg
3.0 to 7.0 ml/kg
GEF
Global Ejection Fraction - equal to 4 times the
stroke volume divided by the global end
diastolic volume (GEDV).
1 to 99%
25 to 35%
PVPI
Pulmonary Vascular Permeability Index -
Indicator of capillary leakage.
0.1 to 10
1.0 to 3.0
CFI
Cardiac Function Index - CFI = CI / GEDVI
1.0 to 15.0 L/min
4.5 - 6.5 L/min
ITBV
Intrathoracic Blood Volume.
50 to 6000 ml
Not applicable
ITBVI
Intrathoracic Blood Volume Index.
ITBVI = ITBV / PBSA (Predicted Body Surface
Area)
100 to
3000 ml/m
2
850 - 1000 ml/m
2
NOTE:
•
Normal values reflect clinical experience and may vary from patient to patient. They are without
guarantee.
• The prefix of “p-” before SV, SVI, SVR, SVRI is an indication that these parameters are derived
by the PiCCO pod.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...