25-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
25 B
ISPECTRAL
I
NDEX
(BIS
X
)
MONITORING
BIS: Sensor expired
The sensor is expired.
Remove the expired sensor and replace
with a new/compatible sensor.
BIS: Check Sensor /
Impedance
• Sensor fault
• Defective sensor cable
• Make sure the sensor is properly
applied to the patient, then perform a
manual impedance check to continue
monitoring. If this fails, do one of the
following:
• Disconnect and examine the sensor.
• Replace the sensor, if necessary.
• Replace the sensor cable.
• Call your hospital’s technical
personnel or DrägerService.
BIS: Pod disconnected
The BISx pod is not
connected to the monitor.
• Check the cables and connections.
• Replace the cables, if necessary.
BIS: Pod Failure
• BISx pod self-test failed
• BISx pod hardware failure
• BISx pod software failure
• Defective sensor
• Check the BISx pod connection.
• Replace the sensor, then disconnect
and reconnect the pod to continue
monitoring.
• Call your hospital’s technical
personnel or DrägerService.
Duplicate devices
connected
Multiple BISx pods or MIB BIS
is connected.
Disconnect the unwanted duplicate pod
or device.
BIS: SRS Final use
SRS cable last use in
progress.
Replace the SRS cable after current
use.
BIS: Replace Sensor
• Unknown sensor type
• Too many uses/sensor worn
out
• Invalid sensor
• Defective sensor cable
• SRS no more uses available
(When SRS is connected
after last use.)
• Remove the unknown/worn out/invalid
sensor and replace with a new/
compatible sensor.
• Replace the sensor cable.
• Replace the SRS cable before
continuing.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...