17-2
D
ELTA
/D
ELTA
XL/K
APPA
VF8
17 P
ULSE
O
XIMETRY
(S
P
O2)
Overview
Dräger offers three algorithms for SpO
2
monitoring:
Masimo SET
®
Nellcor Oximax
®
If installing VF8 software on monitors that do not have a Masimo SET logo
(see page 17-4) or a Nellcor OxiMax logo (see page 17-5), Dräger
OxiSure
(with connection via the MultiMed cables) is available.
These technologies enhance the quality of SpO
2
monitoring, allowing you to measure
the percentage of functional hemoglobin saturated with oxygen (% SpO
2
) in the
patient’s arterial blood. An SpO
2
sensor measures the absorption levels of red and
infrared light. The monitor uses the difference between the two measurements to
calculate the percentage of saturated hemoglobin. Because light absorption varies with
blood volume and blood volume varies with pulse rate, the monitor can also derive a
pulse rate (PLS).
The SpO
2
sensor, available for adult, pediatric, and neonatal patients is connected to
the monitor via the various MultiMed and NeoMed cables, or the Infinity
®
Masimo
SET SpO
2
SmartPod
®
, or the Infinity
®
Nellcor Oximax
®
SpO
2
SmartPod
®
pod.
Precautions
Refer to the “Safety Considerations” section starting on page 6, at the beginning of
these Instructions for Use, for general precautions regarding safe device operation.
SpO
2
measurements are particularly sensitive to the pulsations in the artery and the
arteriole. Measurements may not be accurate if the patient is experiencing shock,
hypothermia, anemia or has received certain medications that reduce the blood flow in
the arteries.
NOTE:
For a complete list of Dräger provided SpO
2
accessories available with this
product,
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...