![Dräger Infinity Delta Instructions For Use Manual Download Page 313](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779313.webp)
M
ICR
O2+® S
TANDALONE
P
ULSE
O
XIMETER
VF8
D
ELTA
/D
ELTA
XL/K
APPA
17-23
®
SpO
2
Status Messages
Message
Possible Cause
Suggested Action
SpO
2
* > #
SpO
2
* < #
The patient’s SpO
2
* is outside
the current upper or lower alarm
limits.
Check the patient and treat, if necessary.
PLS* > #
PLS* < #
The patient’s pulse rate is
outside the current upper or
lower alarm limits.
PLS* Out of Range
(High)
The pulse rate is outside the
measuring range of the monitor
SpO
2
% > UL
The SpO
2
% is greater than
the upper limit.
SpO
2
* Transparent
Too much light is reaching the
sensor’s light detector. This
condition usually occurs
because the sensor is off the
finger or because too much
ambient light is detected.
• Make sure the sensor is properly
attached to the patient’s finger.
• Remove the light source.
• Cover the sensor with opaque material.
• Make sure no ambient light can reach
the detector.
• Contact DrägerService.
SpO
2
*
Unrecognized
Sensor
An invalid value was detected for
the sensor calibration resistor.
• Check for a defective or unapproved
sensor.
• Replace the sensor.
• Contact the hospital’s technical
personnel or DrägerService.
SpO
2
* Light
Blocked
Insufficient light is reaching the
sensor’s light detector.
NOTE:
With detached or partially
detached disposable sensors,
the light emitters and detectors
may have become misaligned.
• Make sure the light sensor is properly
attached to the patient’s finger, and that
the finger is free of blocking substances.
• Check for a defective sensor and
replace, if necessary.
SpO
2
* Artifact
Persistent artifact
• Make sure the SpO
2
sensor is properly
attached to the patient, the monitoring
site is free of patient motion, and all
cables are properly connected.
• Contact DrägerService.
SpO
2
* Weak signal
The pulse amplitude is too low
due to one of the following
conditions:
- Poor perfusion (shock).
- Low body temperature.
• Check the patient.
• Make sure the SpO
2
sensor is properly
attached to the patient and all cables are
properly connected.
• Relocate the sensor to another
extremity.
• Contact DrägerService.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...