A
LARM
S
ETUP
(A
LARM
L
IMITS
T
ABLE
)
VF8
D
ELTA
/D
ELTA
XL/K
APPA
5-11
PLS/PLS*
30 to 300 beats per minute
Off
Adult:
Low
45 bpm
High
120 bpm
Pediatric:
Low
50 bpm
High
150 bpm
Neonatal:
Low
80 bpm
High
180 bpm
PVC/min
Adult and pediatric: 1 to 50 PVC per
minute (upper high limit only)
On
High: 10 PVC per minute
RA
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
Low 2 mmHg (0.3 kPa)
High 12 mmHg (1.6 kPa)
RESP
Adult: 5 to 100 breaths per minute
Pediatric and Neonatal: 5 to 145
breaths per minute
Off
Adult:
Low 5 bpm
High 30 bpm
Pediatric/Neonatal:
Low 20
bpm
High 80 bpm
RRc
5 to 145 breaths per minute
Adult:
Off
Ped/
Neo:
On
Adult:
Low 5 bpm
High 30 bpm
Pediatric/Neonatal:
Low 20
bpm
High 60 bpm
RRc*
5 to 90 breaths per minute
Off
Low 5 bpm
High 30 bpm
RV S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
S:
Low 10 mmHg (1.3 kPa)
High 35 mmHg (4.7 kPa)
M:
Low 7 mmHg (0.9 kPa)
High 17 mmHg (2.3 kPa)
D:
Low 2 mmHg (0.3 kPa)
High 13 mmHg (1.7 kPa)
SpO
2
/SpO
2
2*
20 to 100%
NOTE
: High and low SpO
2
(saturation) alarm conditions are
graded “Medium”.
Adult:
Off
Ped/
Neo:
On
Adult/Pediatric:
Low 90% High 100%
Neonatal:
Low 85% High 95%
SpO
2
%
1 to 100% (high limit only)
Off
Adult/Pediatric:
High 20%
Neonatal:
High 10%
ST Alarms
See “ST Alarms Table” on page 10-10.
Parameter
Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...