ST Alarms Table
The ST Alarms table allows you to modify the alarm limits of multiple ST parameters
in a single location. ST alarms are subject to the same alarm guidelines as other
parameters (see chapter 5). In addition, control keys at the bottom of the screen allow
you to execute the following alarm functions at the ST alarm table:
To access the ST Alarms setup table
1. Press the
Alarm Limits
fixed key on the front of the monitor.
2. Click on the
ST
control key at the lower right of the screen.
3. Follow the guidelines for modifying alarm limits on page 5-6.
4. Use the control keys at the top of the parameter list (
Auto Set
) and bottom of
the screen (
Event Duration
,
Relearn
) to perform other ST alarm functions.
Menu Item
Description
Settings
Auto Set
Changes the upper and lower limits (mm
or mV) for all active ST leads
NOTE
: The main Alarm Limits menu's
Auto Set feature also uses this
calculation to adjust alarm limits of active
ST parameters (see page 5-13).
• Upper alarm limit
Current value + 2mm (or 0.2mV)
• Lower alarm limit
Current value - 2mm (or 0.2mV)
Event Duration
Determines the time that a potential
alarm condition must persist on ST leads
before the monitor classifies it as a valid
alarm condition
• OFF, 15, 30, 45, 60 s
Relearn
Initiates a Relearn of the QRS template
(see page 10-5).
Not applicable
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...