S
TATUS
M
ESSAGES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
17-17
SpO
2
Regulation
Error
• Inconsistent light level
• Excess ambient light
• Make sure the SpO
2
sensor is properly
attached to the patient.
• Remove or cover any external sources
of light entering the sensor.
• Contact DrägerService.
SpO
2
Unplugged
The sensor cable is not
connected to the MultiMed pod.
• Make sure the cables are securely
connected.
• Check for a defective sensor.
SpO
2
: non-Masimo
compatible sensor
A non-Masimo compatible
sensor or SpO
2
pod is
connected to a monitor that is
configured for Masimo.
• Replace with a Masimo-compatible
sensor.
• Contact the biomedical personnel
(SpO
2
compatibility is a locked option).
Duplicate Device
Connected
A MultiMed pod (with SpO
2
sensor) and an Infinity Masimo
SET pod are connected
simultaneously.
Disconnect the duplicate device.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...