OR A
LARMS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
5-15
OR Alarms
Monitoring may be interrupted or discontinued more frequently during anesthesia
than in the course of critical care. For this reason, some alarms behave differently
when the monitor operates in OR mode. As shown in the following table, certain
alarms stop annunciating when the condition ceases (one-shot alarms), while others
emit a single attention tone. For more information on OR mode, see page 2-12.
NOTE:
For anesthesia alarms you cannot set the menu selection
Attention Tone
Volume
to
OFF
.
Message
Condition
Priority
Annunciation
SpO2 Transparent
Nothing detected between the sensor’s
light source and the detector
Low
One-shot
SpO2 Light Blocked
Insufficient light for valid measurement
Low
One-shot
ECG Leads Invalid
• QRS processing leads invalid for > 10
seconds
• Faulty electrode contact or lead set
• Unplugged lead set
• Wrong Cable Type selected on ECG
Lead Setup menu (see page 8-18)
Low
One-shot
Apnea
Breath has not been detected for Apnea
Time (
A
T
)
seconds
No
alarm
Single
attention tone
Breath has not been detected for
A
T
x 2 seconds
No
alarm
Single
attention tone
Breath has not been detected for
A
T
x
3 seconds
Medium
One-shot
Breath has not been detected for
A
T
x
6 seconds
High
One-shot
NOTES
:
• The information in this table applies to apnea detected by etCO
2
monitoring from any source
(etCO
2
pod or module, MultiGas module, etc.). You can set RRc Apnea Time (A
T
) to OFF, 10,
15, 20, 25 or 30 seconds on setup menus for etCO
2
or etCO
2
*.
• If you press the
Alarm Silence
key any time after the monitor’s first indication of an apnea
condition (A
T
), subsequent alarms for that apnea condition do not annunciate.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...