Overview
The Infinity PiCCO pod monitors cardiopulmonary and circulatory variables. Cardiac
output is determined both continuously through arterial pulse contour analysis and
intermittently through transpulmonary thermodilution technique. In addition, the
PiCCO pod measures systolic, diastolic and mean arterial pressure. Analysis of the
thermodilution curve is used to determine the intra- and extra- vascular fluid volumes.
If a patient’s weight and height are entered, the PiCCO pod indexes the derived
parameters to body surface area (BSA) and body weight (BW) respectively. The
PiCCO pod monitors adult and pediatric patients (it is not available for neonatal
patients).
The pulse contour PiCCO parameters are listed in the following table.
Pulse Contour parameters
Label
Parameter
Measurement
Range
Normal Value
PCCO
Pulse Contour Cardiac Output - uses a
calibration factor (cal) determined by the
thermodilution cardiac output
measurement and the heart rate, as well as
the integrated values for the area under the
systolic part of the pressure curve, the
aortic compliance and the shape of the
pressure curve, represented by change of
pressure over change of time.
0.25 to 25 L/min
Not applicable
PCCI
Continuous Cardiac Index - Pulse Contour
Cardiac Output normalized to body surface
area.
0.01 to 15
L/min/m
2
3.00 to 5.00
L/min/m
2
p-SV
Stroke Volume
1 to 250 ml
Not applicable
p-SVI
Stroke Volume Index - Stroke Volume
normalized to body surface area.
1 to 125 ml/m
2
40 to 60 ml/m
2
p-SVR
Systemic Vascular Resistance - The
quotient of driving pressure and cardiac
output over the last 12 seconds. Here the
driving pressure represents the difference
between mean arterial pressure (MAP) and
central venous pressure (CVP).
0 to 30000
dynes x sec x cm
-5
Not applicable
NOTE:
• Normal values reflect clinical experience and may vary from patient to patient. They are without
guarantee.
• The prefix of “p-” before SV, SVI, SVR, SVRI is an indication that these parameters are derived
by the PiCCO pod.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...