Hemodynamic Pods
User Interface
User Controls:
Fixed keys (C.O. Start, IBP Zero, Wedge)
Displays:
HemoMed:No LCDs
Hemo2: two 4-character LCDs.
Hemo4: four 4-character LCDs.
MPod – Quad Hemo: No LCDs
Connections:
Hemo2: 2 invasive pressures, C.O., 2 temperatures, single
cable connecting pod to monitor
Hemo4, MPod – Quad Hemo, and HemoMed:
4 invasive pressures, C.O., 2 temperatures, single cable
connecting pod to monitor
Degree of protection against ingress
of water:
Hemo2, Hemo4, and HemoMed: IPX0 (per IEC 60529)
MPod – QuadHemo: IPX1 (protected against harmful
effects of dripping water) per IEC 60529
Physical Attributes
Size (H x W x D):
Hemo2, Hemo4, and HemoMed:140 x 205 x 60 mm
(5.5 x 8.1 x 2.3 in)
MPod – QuadHemo: 205 x 110 x 80 mm (8.1 x 4.3 x 3.2 in)
Weight:
Hemo2 and HemoMed:0.7 kg (1.6 lb.)
Hemo4: 0.9 kg (1.9 lb.)
MPod – Quad Hemo: 0.48 kg (1.1 lb)
NOTE:
Hemo2 weight includes one transducer adapter
block. Hemo4 weight includes two transducer adapter
blocks. Both weights exclude mounting clamp.
For the MPod – Quad Hemo the weight includes four
transducer cables but excludes the mounting clamp and
rod.
Electrical Specifications
Power source:
Powered directly from the monitor
Protection against electric shock:
Type CF
Mode of operation:
Continuous
Defibrillation protection:
Per IEC 60601-2-34
Environmental Requirements
Temperature range:
Operating:
Hemo2, Hemo4, HemoMed:10 to 40 °C (50 to 104 °F)
MPod – QuadHemo: 0°C to 45°C (32°F to 113°F)
Storage:
Hemo2, Hemo4, HemoMed: –20 to 50 °C (–4 to 122 °F)
MPod – QuadHemo: –40°C to +70°C (–40°F to +158°F)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...