29-8
D
ELTA
/D
ELTA
XL/K
APPA
VF8
29 C
LEANING
AND
D
ISINFECTING
Airway Adapter
1. Rinse airway adapters in a warm soapy solution, then soak them in a liquid
disinfectant.
2. Dry with a lint-free cloth, making sure adapter windows are dry and free of
any residue before they are used.
Sidestream Sampling Pump (etCO
2
only)
The etCO
2
module and pod contain a small pump that draws air from the nasal cannula,
through the sidestream airway adapter, and out the exhaust port. Suggested cleaning proce-
dures are outlined below.
The following fluids are acceptable for cleaning:
Isopropyl alcohol.
A 5.25 % water solution (by weight) of sodium hypochlorite (bleach).
The following items are required for cleaning the etCO
2
module or pod:
A 60 cc catheter-tip syringe.
A 2-foot section of 1/8- or 3/16-inch tubing to drain off fluid after it passes
through the etCO
2
pump.
A receptacle to receive the fluid after it drains.
CAUTION:
To reduce the risk of infection, remember that airway and flow sensors are for
single patient use only and cannot be sterilized. To avoid blocking the sampling
tube, do not attempt to sterilize the pod by immersing it in liquids.
Do not attempt to sterilize the etCO
2
/Respiratory Mechanics pod by immersing
it in liquids.
WARNING: Used sampling tubing, T-connectors, and water
traps could be contaminated and must be handled and
disposed of with care. Infection hazard may be present.
Dispose of these items in accordance with local regulations.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...