5-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
5 A
LARMS
LV S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
S:
Low 75 mmHg (10.0 kPa)
High 160 mmHg (21.3 kPa)
M:
Low 40 mmHg (5.3 kPa)
High 80 mmHg (10.7 kPa)
D:
Low 2 mmHg (0.3 kPa)
High 25 mmHg (3.3 kPa)
NBP S/M/D
Adult: 10 to 250 mmHg
(1.3 to 33.3 kPa)
Pediatric: 10 to 170 mmHg
(1.3 to 22.7 kPa)
Neonatal: 10 to 130 mmHg
(1.3 to 17.3 kPa)
Off
S: (Adult)
Low 90 mmHg (12.0 kPa)
High 160 mmHg (21.3 kPa)
M: (Adult)
Low 60 mmHg (8.0 kPa)
High 125 mmHg (16.7 kPa)
D: (Adult)
Low 50 mmHg (6.7 kPa)
High 110 mmHg (14.7 kPa)
S: (Pediatric)
Low 50 mmHg (6.7 kPa)
High 120 mmHg (16.0 kPa)
M: (Pediatric)
Low 40 mmHg (5.3 kPa)
High 85 mmHg (11.3 kPa)
D: (Pediatric)
Low 35 mmHg (4.7 kPa)
High 80 mmHg (10.7 kPa)
S: (Neonatal)
Low 50 mmHg (6.7 kPa)
High 80 mmHg (10.7 kPa)
M: (Neonatal)
Low 40 mmHg (5.3 kPa)
High 70 mmHg (9.3 kPa)
D: (Neonatal)
Low 25 mmHg (3.3 kPa)
High 60 mmHg (8.0 kPa)
PA S/M/D
-5 to 300 mmHg (-0.7 to 40.0 kPa)
Off
S:
Low 10 mmHg (1.3 kPa)
High 35 mmHg (4.7 kPa)
M:
Low 7 mmHg (0.9 kPa)
High 17 mmHg (2.3 kPa)
D:
Low 2 mmHg (0.3 kPa)
High 13 mmHg (1.7 kPa)
Parameter
Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...